The radiant panel subassembly has been completed with the
exception of the support brackets and reverberatory screen. The equipment
can be assembled to permit test of specimens of thickness up to 50
mm.
5.1 The panel frame should be placed upright on
a level floor, preferably in the location in which the equipment will
be used.
5.2 The rotating ring should be mounted on its
three guide bearings.
5.3 The panel mount frame should be bolted together,
and to the ring, by four bolts.
5.4 A check should be made that the ring lies
in a vertical plane. If the error is large, an adjustment of the upper
ring support-bearing location may be necessary. Prior to making such
an adjustment, it should be determined whether the error is due to
excessive clearance between the ring and bearing rollers. If this
is the case, rollers of a larger diameter may correct the problem.
5.5 The four panel support brackets should be
fastened to the radiant panel at the four corners. Do not use too
much force in bolting these brackets in place. Prior to mounting these
brackets, one 35 mm M9 cap screw is placed in the hole that will be
farthest from the panel end. These screws provide a means for mounting
the panel.
5.6 Four washers should be placed on each of the
panel mounting screws and the panel assembled on the mount bracket.
5.7 The angularity of the radiant panel surface
with the plane of the mounting ring should be checked. This can be
accomplished by means of a carpenter's square and measurements to
the refractory tile surface at both ends of the panel. Any deviation
from the required 15° angle may be adjusted by increasing or reducing
the number of washers on the mounting screws.
5.8 The radiant panel should be rotated to face
a specimen mounted in a vertical plane.
5.9 The panel surface should be checked with a
spirit level to ensure that it also lies in a vertical plane.
5.10 The specimen frame with specimen support
rails on the side and bottom positions and the pilot burner holder
assembled in approximate positions should be brought up to the burner
frame and the two frames fastened together with two bolts and six
nuts or two threaded rods and eight nuts. The spacing between the
frames is roughly 125 mm.
5.11 The spacing of the two sides of the frames
is adjusted to ensure that the specimen support frame longitudinal
members are at a 15° angle to the radiant panel surface.
5.12 The single specimen holder side guide rail
for vertical specimen orientation should be adjusted so that it is
at the required 15° angle to the radiant panel surface.
5.13 An empty specimen holder should be slid into
position on the rail and the position of the upper guide fork adjusted
to ensure that when a specimen is inserted in the holder its surface
will lie in a vertical plane.
5.14 The stop screw determining the axial position
of the specimen holder should be adjusted to ensure that the axis
of the pilot burner is 10 ± 2 mm from the closest exposed edge
of the specimen. This adjustment should again be made by use of an
empty specimen holder and substitution of a 6 mm steel rod of 250
mm length for the pilot burner ceramic tube. When viewed from the
back of the specimen holder, the spacing between rod axis and the
edge of the specimen retaining flange of the holder should be 10 ±
2 mm.
5.15 With the specimen holder still in place against
the stop screw, the spacing between the panel and specimen support
frames should be adjusted to make dimension B (see figure 3) equal
to approximately 125 mm. This adjustment is made by means of the two
screws fastening the frames together. In making this adjustment, it
is important to make equal adjustments on each side to maintain the
angular relationship called for in adjustments in paragraphs 5.11
and 5.12.
5.16 The nuts supporting the specimen holder side
guide rail should be adjusted to ensure that dimension A (see figure
3) is 125 ± 2 mm. Again, equal adjustments to the two mounting
points are required. When doing this, a check should be made to ensure
that the guide rail and edge of the specimen holder are in a horizontal
plane. In making this adjustment, it is important to ensure that the
45 mm stack position dimension, as shown in figure 4, is maintained.
Another way of adjustment to dimension A is through changes in the
number of washers mentioned in paragraph 5.6.
5.17 If necessary, the procedure described in
paragraph 5.13 should be repeated.
5.18 The reverberatory screen should be mounted
on the radiant panel. This must be done in such a manner that it is
free to expand as it heats up during operation.
5.19 The viewing rake with 50 mm pins is mounted
on an angle fastened to the specimen holder guide rail. Its position
is adjusted so that pins are located at multiples of 50 mm distance
from the closest end of the specimen exposed to the panel. It should
be clamped in this position.
Table 1 – Calibration of incident heat flux to the specimen
|
Distance from exposed end of the specimen (mm)
|
Typical flux levels at the specimen (kW/m2)
|
Calibration position to be used (kW/m2)
|
0
|
49.5
|
|
50
|
50.5
|
50.5
|
100
|
49.5
|
|
150
|
47.1
|
x
|
200
|
43.1
|
|
250
|
37.8
|
x
|
300
|
30.9
|
|
350
|
23.9
|
23.9
|
400
|
18.2
|
|
450
|
13.2
|
x
|
500
|
9.2
|
|
550
|
6.2
|
x
|
600
|
4.3
|
|
650
|
3.1
|
x
|
700
|
2.2
|
|
750
|
1.5
|
x
|
Typical flux incident on the specimen and specimen positions
at which the calibration measurements are to be made. The flux at
the 50 mm and 350 mm positions should agree with the typical values
within 5%. Calibration data at other positions should agree with typical
values within 10%.
Two sets of thermocouples (T.C.) and lead wires are required.
The wire size and lengths within the fume T.C. group must be the same
to ensure proper signal averaging. The parallel connection of the
couples may be achieved at the mixing box by plug connection of the
leads. This allows quick removal and checks for continuity and grounding
problems with minimum delay. No cold junction should be used but the
signal mixing box should be shielded from panel radiation.