3.1 Additional instrumentation consists of a restricted-view
total-heat flux meter calibrated with the restricted view to indicate
incident heat flux. The flux meter shall be water-cooled and capable
of measuring heat flux 0 kW/m2 to 60 kW/m2.
The flux meter should be calibrated at least once a year against a
standard device.
3.2 The flux meter should be placed perpendicular
to the centre of the window being tested, and in a position such that
the centre of the flux meter's view coincides with the centre of the
windowfootnote (see figure). The flux meter should
be located at a distance greater than 0.5 m from the window, such
that the view of the flux meter just includes part of the frame. However,
the flux meter should not be located more than 2.5 m from the window.
The dimension of the boundary and window frame seen by the flux meter,
which remains outside the window, should not exceed 10% of the total
width seen by the flux meter on the surface of the sample. It should
be calculated on the basis of restricted view angle of the flux meter
and its distance to the sample surface.
3.3 For windows whose greater dimension is less
than 1.57 times the smaller dimension, only one flux meter is needed.
3.4 For oblong windows whose greater dimension
is more than 1.57 times the smaller dimension, additional flux meters
should be provided. The distance of the flux meters from the window
should be adjusted such that the flux meters' view covers at least
50% of the window. However, the flux meters should not be located
less than 0.5 m nor more than 2.5 m from the window.